13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
250
13.2.1.7.2
Manual adjustment: Ceph secondary diaphragm (QuickShot)
The manual adjustment procedure is similar to the one for automatic
adjustment [ → 247]. The only difference is that the default adjustment
values calculated automatically in the
diaphragm (Quickshot)"
calculated manually.
1. Start the
adjustment procedure as described in the chapter entitled "Automatic
adjustment: Ceph - QuickShot fixed point of rotation
(standard) [ → 247]".
"Adjustment of Ceph secondary
menu are overwritten by adjustment values
"Adjustment of Ceph secondary diaphragm (Quickshot)"
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016