Sensor Calibration - Sirona ORTHOPHOS SL 2D Installation Manual

Hide thumbs Also See for ORTHOPHOS SL 2D:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016

Sensor calibration

13.2.2.3
1. In the structure tree, under
element (S010.64).
2. Establish [ → 209] receptivity.
3. Create an exposure (85 kV / 22 mA s) [ → 209].
 A message is displayed in the action window to indicate that the
exposure is complete and the status indicator in front of the
"Sensor calibration"
accordingly:
Calibration not OK = status indicator stays red
Calibration OK = status indicator turns green
NOTE: The calibration values are automatically saved.
4. Continue with the next stage of the calibration procedure [ → 266].
13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
"3D"
, click on the
element in the structure tree responds
"Sensor calibration"
265

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos sl 2d / cephOrthophos sl 3dOrthophos sl 3d / ceph

Table of Contents