Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
13.2 Adjustment and calibration via the calibration menu
1. In the structure tree, under
secondary diaphragm"
"Adjustment of Ceph secondary diaphragm"
The
displayed in the action area. The menu supports precision
adjustment and coarse adjustment (precision adjustment is pre-
selected). Perform a precision adjustment first. In most cases,
previous coarse adjustment is not necessary. [ → 203]
2. Establish readiness for exposure [ → 209].
3. Create an exposure (80kV / 14 mA; 0.6 s) [ → 209].
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
"Adjustment of Ceph secondary diaphragm"
the
The exposure is displayed in the exposure window.
13 Unit adjustment and calibration
"Ceph"
"Adjustment of Ceph
, click on the
element (S010.5).
menu is
menu.
239