13.2 Adjustment and calibration via the calibration menu
1. In the structure tree, under
secondary diaphragm"
"Adjustment of Ceph secondary diaphragm"
 The
displayed in the action area. The menu supports precision
adjustment and coarse adjustment (precision adjustment is pre-
selected). Perform a precision adjustment first. In most cases,
previous coarse adjustment is not necessary. [ → 203]
2. Establish readiness for exposure [ → 209].
3. Create an exposure (80kV / 14 mA; 0.6 s) [ → 209].
 The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
"Adjustment of Ceph secondary diaphragm"
the
 The exposure is displayed in the exposure window.
13 Unit adjustment and calibration
"Ceph"
"Adjustment of Ceph
, click on the
element (S010.5).
menu is
menu.
239