14 Final work
14.1 Save sensor data
"Backup / Restore sensor data"
Open the
276
14

Final work

14.1

Save sensor data

You can use the
Backing up sensor data1
(factory calibration of the sensor) from the DX83 board and save it as a
Zip file.
NOTE: Since the sensor data is saved on the DX83 board, this is required
e.g. before exchanging the DX83 board. The saved sensor data must
then be resaved on the new DX83 board following replacement of the
board. This is possible via the
Service Manual).
The unit and computer are turned on.
Open the backup / restore data menu
The unit is logged into the network as an X-ray component.
1. Start an Internet browser such as Internet Explorer or Firefox.
2. In the address line, enter "https://" and the IP address of your unit.
Example: https://192.168.15.240
NOTICE! The IP address of your unit can be found on the Info screen.
3. Confirm your input with the "Enter" key.
 A security prompt appears.
4. Click on the link
 The Sirona Web service is loaded.
 The sub menus of the
tree.
Installation Manual ORTHOPHOS SL
"Backup sensor data"
menu to load the sensor data
"Restore sensor data"
"Continue to this Website"
"User"
menu are displayed in the structure
Sirona Dental Systems GmbH
menu (see the
.
64 95 142 D3632
D3632.031.01.02.02 03.2016