Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
Geometry phantom for volume calibration
13.1.6.3
NOTICE
Risk of damage to unit
Make absolutely sure that the geometry phantom is inserted in the bite
block holder of the unit in such a way that the body of the phantom and
the screw (C) point
figure). Otherwise, this can result in a collision between the sensor unit
and the geometry phantom and thereby damage the unit.
* For greater clarity, the unit is shown in the left-hand image without
sensor.
"Sensor adjustment"
For
*
calibration, you must insert the geometry phantom (A) into the pan bite
block holder (B) on the unit.
"Diaphragm adjustment"
For
calibration (DCS)"
must take the geometry phantom back out of the bite block holder.
13 Unit adjustment and calibration
13.1 General information on unit adjustment and calibration
away from the unit in the direction of the patient
3D adjustment and
3D adjustment, as well as for
"Diaphragm calibration"
and
(see
"Geometry calibration"
3D
"Sensor
3D calibrations, you
A
C
B
213