13 Unit adjustment and calibration
13.2 Adjustment and calibration via the calibration menu
216
13.2.1.1.1
Automatic adjustment: Panoramic sensor (standard)
✔
The needle phantom is in the bite block holder on the unit [ → 210].
1. Open the calibration menu [ → 195].
2. At the bottom of the structure tree, click on the
element (S010.1) under
The
The menu supports precision adjustment and coarse adjustment
(precision adjustment is pre-selected). Perform a precision
adjustment first. In most cases, previous coarse adjustment is not
necessary. [ → 203]
3. Establish receptivity [ → 209].
4. Create an exposure (60 kV / 3 mA ; 0.7 s). [ → 209]
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
the
The exposure is displayed in the exposure window.
"Pan"
.
"Sensor adjustment"
menu is displayed in the action area.
"Sensor adjustment"
menu.