Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016
Pan symmetry adjustment
13.2.1.3
13.2.1.3.1
Automatic adjustment: Pan - Symmetry
✔
The needle phantom is in the bite block holder on the unit [ → 210].
1. In the structure tree, under
adjustment"
The
2. Establish [ → 209] receptivity.
3. Create an exposure (60 kV / 3 mA; 14.1 s) [ → 209].
The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
the
The exposure is displayed in the exposure window.
13.2 Adjustment and calibration via the calibration menu
"Pan"
element (S010.2).
"Pan - Symmetrie"
menu is displayed in the action area.
"Symmetrie adjustment"
13 Unit adjustment and calibration
"Symmetrie
, click on the
menu.
227