Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
64 95 142 D3632
D3632.031.01.02.02 03.2016

Pan symmetry adjustment

13.2.1.3
13.2.1.3.1
Automatic adjustment: Pan - Symmetry
The needle phantom is in the bite block holder on the unit [ → 210].
1. In the structure tree, under
adjustment"
 The
2. Establish [ → 209] receptivity.
3. Create an exposure (60 kV / 3 mA; 14.1 s) [ → 209].
 The adjustment values S1, S2 and S3 are automatically
determined from the exposure and entered in the text boxes of
the
 The exposure is displayed in the exposure window.
13.2 Adjustment and calibration via the calibration menu
"Pan"
element (S010.2).
"Pan - Symmetrie"
menu is displayed in the action area.
"Symmetrie adjustment"
13 Unit adjustment and calibration
"Symmetrie
, click on the
menu.