64 95 142 D3632
D3632.031.01.02.02 03.2016
S018: Test step 8
11.12.4.7
Functional test of the occlusal bite block
Symbol on the control
panel
Patient symbol key 1 (G) visible
Selection field
1
2
3
All of the monitoring functions of the bite block electronics are active in
this service routine. Any error states are displayed as a code in selection
field 3.
1. Call service routine S018.8.
 The current angle of the bite block plate or the one calculated
based on the calibration is displayed in selection field 1.
 The currently measured AD value leading to the angle calculation
is displayed in selection field 2.
 The codes of the possible error modes of the acquisition
electronics are displayed in selection field 3 (see section "Codes
of possible error modes [ → 156]").
 The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
2. Exit the service routine.
11 Commissioning, demo mode and repacking
11.12 Service routines for commissioning and repacking
Status
Meaning
Occlusal bite block
connected
not visible
Occlusal bite block not
connected
Display
Angle of the bite block plate calculated on the basis of
the calibration
Measured AD value providing the basis for the angle
calculation
Codes for possible error modes [ → 156]
155