Concluding the measurement
64 95 142 D3632
D3632.031.01.02.02 03.2016
1. Switch the unit on via switch (A) (see also Operating Instructions).
2. Loosen the screw and remove the lid of the tube assembly cover.
3. DANGER! After switching off the unit, wait at least 4 minutes before
removing the measuring wires or reinserting the jumper!
Remove the measuring wires and bridge with the test points MA+/
MA- on the DX6 board again with the jumper (H).
4. Reattach the cover plate (G) to the tube assembly with the 4 screws
(F).
F
5. Reattach the lid of the tube assembly cover to the unit and secure it
with the screw.
G
12 Startup, measurements and tests (for USA/Canada only)