6 Installing the sensor unit
6.2 Installing the sensor cover on the ring
66
6.2
Installing the sensor cover on the ring
1. Loosen screw (G) (approx. 2-3 turns).
2. Insert the sensor cover (H) from below over the sensor bracket and
then slide it under the ring cover.
3. Screw the sensor cover onto the rotation unit with the 2 screws (J).
4. Tighten the screw (G) firmly.
Sirona Dental Systems GmbH
Installation Manual ORTHOPHOS SL
G
J
H
64 95 142 D3632
D3632.031.01.02.02 03.2016