13 Switch to the power sensor's lower range:
• Press
• Move to the Range parameter using the up/down arrow keys.
• Scroll through the Range values by pressing
LOWER.
• Press
14 Disconnect the 30 dB attenuator and reconnect the test setup as shown in
Figure 2-
15 Turn on the power reference:
Press
16 Verify that an overload error is displayed on the status line at the top of the
power meter's display and that the display reads 999.99 dBm.
17 Switch to the power sensor's upper range:
• Press
• Move to the Range parameter using the up/down arrow keys.
• Scroll through the Range values by pressing
UPPER.
• Press
18 Verify that the display reads 0 dBm ±1 dB. Record the reading.

Table 2-6 Functional Test (Average Path) Results

Low Range, 30 dB pad
Upper Range, 30 dB pad
46
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Channel
.
Done
.
7.
Zero
,
Power Ref On
Cal
.
Channel
.
Done
Test
_________________
_________________
Low Range, no pad
_________________
High Range, no pad
_________________
Change
. Select a value of
Change
. Select a value of
Channel A Results
Channel B Results (Agilent
_________________
_________________
_________________
_________________
E4416A/E4417A Power Meters Service Guide
E4417A Only)