Abort Commands - Keysight Technologies N6705C Operating And Service Manual

Dc power analyzer
Table of Contents

Advertisement

7 SCPI Programming Reference

Abort Commands

Abort commands cancel any pending triggered actions. They return the specified trigger system back
to the Idle state.
ABORt:ACQuire (@<chanlist>)
ABORt:DLOG
ABORt:ELOG (@<chanlist>)
ABORt:HISTogram (@<chanlist>)
ABORt:TRANsient (@<chanlist>)
ABORt:ACQuire - Cancels any triggered measurements. It also resets the WTG-meas bit in the
Condition Status register.
ABORt:DLOG - Stops internal data logging.
ABORt:ELOG - Stops external data logging.
ABORt:HISTogram - Stops the histogram current measurement.
ABORt:TRANsient - Cancels any triggered actions. It also resets the WTG-tran bit in the Operation
Condition Status register.
Parameter
(none)
Cancel the measurement: ABOR:ACQ (@1)
Stop the internal data log: ABOR:DLOG
Stop the external data log: ABOR:ELOG (@1)
Cancel the histogram measurement: ABOR:HIST (@1)
Cancel the triggered action: ABOR:TRAN (@1)
ABORt:ACQuire and ABORt:TRANsient are also executed at power-on and upon execution of the
l
*RST
command.
Abort:TRANsient does not turn off continuous triggers if
l
been programmed. In this case, the trigger system will automatically re-initiate.
232
Typical Return
(none)
INITiate:CONTinuous:TRANsient ON
Keysight N6705C Operating and Service Guide
has

Advertisement