d
To W ait for Sweep End
c
Enable Sweep-End Bit
Line 60 clears all bits of the Status Registers and the Enable Registers.
In line 70, the command
enables bit 2 of the Status Byte Register). The command
Register B to 0000000000000001 (this enables bit 0 of the Event Status Register B. See
Figure 3-4).
Figure 3-3. Sample Program : To Wait for Sweep End
sets the Service Request Enable Register to 00000100 (this
Figure 3-4. Sweep-End Bit Enabling
sets the Event Status Enable
Synchronizing the Analyzer from Remote
a
b
3-3