Response And Isolation Error Correction For Reflection Measurements - Agilent Technologies 8753ET User Manual

Network
Hide thumbs Also See for 8753ET:
Table of Contents

Advertisement

NOTE
You can save or store the measurement correction to use for later measurements. Refer to
Chapter 4 , "Printing, Plotting, and Saving Measurement Results"
14. This completes the response and isolation correction for transmission measurements. You can connect
and measure your device under test.

Response and Isolation Error Correction for Reflection Measurements

The response and isolation error correction for reflection measurements provides the following benefits:
• removes frequency response of the test setup
• removes isolation in transmission measurements
• removes directivity in reflection measurements
To perform the response and isolation error correction for reflection measurements:
1. Press
.
Preset
2. Select the type of measurement you want to make.
❏ If you want to make a reflection measurement on PORT 1 (in the forward direction, S
instrument default setting.
❏ For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the reverse direction,
S
), press:
22
Meas
Refl: REV S22 (B/R)
3. Set any other measurement parameters that you want for the device measurement: power, sweep type,
number of points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
5. If your calibration kit is different than the kit specified under the
CAL KIT
SELECT CAL KIT
If your type of calibration kit is not listed in the displayed menu, refer to
page
7-57.
6. To select a response and isolation correction and to start with the response portion of the calibration,
press:
CALIBRATE MENU
7. Connect the short or open calibration standard to the port you selected for the test port (PORT 1 for S
or PORT 2 for S
on ES analyzers).
22
NOTE
Include any adapters that you will have in the device measurement. That is, connect the
standard device to the particular connector where you will connect your device under test.
Frequency Response and Isolation Error Corrections
(select your type of kit)
RESPONSE & ISOL'N
Calibrating for Increased Measurement Accuracy
CAL KIT [ ]
RETURN
"Modifying Calibration Kits" on
RESPONSE
for procedures.
), leave the
11
softkey, press:
11
6- 19