Response And Isolation Error Correction For Reflection Measurements - Agilent Technologies 8719ES User Manual

Network analyzer
Hide thumbs Also See for 8719ES:
Table of Contents

Advertisement

12.Return the averaging to the original state of the measurement. For example, reduce the
averaging factor by at least four times or turn averaging off.
13.To compute the isolation error coefficients, press:
RESUME CAL SEQUENCE
Cal
The analyzer displays the corrected data trace. The analyzer also shows the notation
Cor at the left of the screen, indicating that the correction is switched on for this
channel.
You can save or store the measurement correction to use for later
NOTE
measurements. Refer to
Measurement Results"
14.This completes the response and isolation correction for transmission measurements.
You can connect and measure your device under test.
Response and Isolation Error Correction for Reflection
Measurements
The response and isolation error correction for reflection measurements provides the
following benefits:
• removes frequency response of the test setup
• removes isolation in transmission measurements
• removes directivity in reflection measurements
To perform the response and isolation error correction for reflection measurements:
1. Press
.
Preset
2. Select the type of measurement you want to make.
If you want to make a reflection measurement on PORT 1 (in the forward direction,
S
), leave the instrument default setting.
11
For ES analyzers, if you want to make a reflection measurement on PORT 2 (in the
reverse direction, S
Refl: REV S22 (B/R)
Meas
3. Set any other measurement parameters that you want for the device measurement:
power, sweep type, number of points, IF bandwidth.
4. To access the measurement correction menus, press:
Cal
DONE
Chapter 4 , "Printing, Plotting, and Saving
for procedures.
), press:
22
Calibrating for Increased Measurement Accuracy
Frequency Response and Isolation Error Corrections
RESP ISOL'N CAL
6- 19

Hide quick links: