Operator Self Tests
.-
Introduction
This appendix gives you an overview of the self tests the logic analyzer
runs when you turn it on. You can also access the self tests from the I/O
menu. This appendix is not intended to provide service information,
but to acquaint you with the tests. If service is required, it should be
performed by qualified service personnel.
Self Tests
The power-up self test is a set of tests that are automatically performed
when you apply power to the logic analyzer. You may perform the self
tests individually to have a higher level of confidence that the
instrument is operating properly. A message that the instrument has
failed a test will appear if any problem is encountered during a test.
The individual self tests are listed in the self test menu which is
accessed via the I/O menu. The HP 1652B/1653B self tests are on the
operating system disk and the disk is required to run the tests.
POWer-Up
Self
The power-up self test is automatically initiated at power-up by the
Test
HP 1652B/1653B Logic Analyzers. The revision number of the
operating system firmware is given in the upper right of the screen
during the power-up self test. As each test is completed, either "passed"
or "failed" will be displayed before the test name as shown below.
PERFORMING POWER-UP SELF TESTS
passed ROM test
passed RAM test
passed Interrupt test
passed Display test
passed Keyboard test
passed Acquisition test
passed Threshold test
passed Disk test
LOADING SYSTEM FILE