Software
Calibration
Procedures
I
Note @
Offset Calibration
I
Note @
The following calibration procedures should be performed in their
entirety and in the same sequence shown in this procedure. The
following test equipment is recommended:
l
BNC cable.
l
DC Power Supply HP 6114A (2
0.1%
accuracy).
l
Digital Voltmeter HP 3478A (2 0.025 % accuracy).
l
BNC (female)-to-dual Banana Adapter.
l
BNC -to-mini probe adapter
An instrument warm-up of 15 minutes is recommended before starting
these procedures. To abort any calibration procedure, use the
front-panel knob to select the Cancel field, then press the
SELECT
key.
1.
In the System Configuration menu turn both State/Timing
analyzers Off, and turn the oscilloscope On.
2. Press the
TRACE/TRIG
key and select the Calibration field using
the front-panel knob and SELECT key.
Offset calibration should be the default Calibration menu setting. If
not, select the Calibration choice field and, when the pop-up appears,
select Offset.
3. Disconnect all signals from the channel 1 and 2 inputs. Select
Start with the front-panel knob and SELECT key. A message will
appear on screen to indicate the calibration is in process.
4.
When the calibration is complete, the calibration status screen will
appear.
Installation, Maintenance
and Calibration
HP 16528/16538
D-12