Clock, Qualifier, And Data Inputs Test 1; Test Connector - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Clock, Qualifier,
Description:
and Data Inputs
Test 1
This test verifies the setup and hold times for the falling edge of the HP 1652B
L clock specification, and the falling edge of the HP 1653B J and K clock
specification.
This test also verifies the maximum clock rate with counting mode
on.
Specification:
Clock repetition rate: With time or state counting mode on, minimum
time
between states is 60 ns.
Setup time: Data must be present prior to the clock transition,
> 10 ns.
*HP 1652B hold time: Data must be present after the falling edge of the L clock
transition, 0 ns.
HP 1653B hold time: Data must be present after the falling edge of the J and K
clock transition, 0 ns.
Equipment
Required:
Pulse Generator
..........................
HP
8161 A/O20
Oscilloscope
............................
HP
54502A
50
Ohm Feedthrough
(2)
...................
HP
IOlOOC
Test Connector
(2)
........................
see figure
3-l
and
3-2
BNC m-m Coupler
(2)
.....................
HP
1250-0216
BNC Cable
(2)
...........................
HP
10503A
BNCTeem-f-f
(2) .........................
HP1250-0781
Procedure:
1. Connect the HP 1652B/1653B and test equipment as in figure 3-3.
OSCILLOSCOPE
PULSE
GENERATOR
LOGIC
ANALYZER
CLK
BIT
1
P
E
DATA
BITS
2
D
1
/
/
1
8
FIGURE
3-l
,
GND
01650849
Figure
3-3.
Setup for Data Test 1
--
Note
11 I s
In this setup, only eight channels are tested at one time to minimize loading. The
ground lead must be connected to ensure accurate test results.
.
_-.
Performance
Tests
HP
1652B/1653B
3-6
Service Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

1653b

Table of Contents