4. Unassign all of the pods frgm the analyzers as shown in figure 3-44. Refer to
Performance
Tests
3-38
steps a through c if you are unfamiliar with menus.
I
System
Configuration
Clnalyter
2
Type:
[o"]
Figure
3-44.
System Configuration
a. Move the cursor to an assigned pod and press SELECT.
b. Move the cursor to
Unassigned
and press SELECT.
c. Repeat steps a and b for all assigned pods.
5. Press the FORMAT/CHAN
key and turn off channel
2
by deleting the
channel 2 waveform as in figure 3-45. Refer to steps a and b if you are
unfamiliar with menus.
II)-
Channel
Input
Probe
s/Dlv
V/Div
[ 1.500
V )
Impedance
[]
Delay
I
Offset
Preset
pZEFTJ
I
rCH1
1
1
1
1
1
I
1
I
I
I
:
:
:
:
:
:
.
.
:
:
:
.
: .
.
:
:
:
.
:
:
.
:
,
t
I
I
I
,
t
1
I
I
I
Figure
3-45.
Deleting
Channel
2
a. Move the cursor to
CH
2 at the left side of the display and press
SELECT.
b. Move the cursor to
Delete waveform
and press SELECT.
.-
HP 1652B/1653B
Service Manual