2. Adjust the pulse generator for the output in figure
3-4.
I+--
60NS
---WI
1
----
3.2v
DATA,OUTPUT
A
+
10NS
CLOCK
OUTPUT
B
L
----
Ov
0165CM'lO
Figure
3-4.
Waveform
for Data Test
1
Setting for HP
8161A:
Parameter
Output A
Output B
Input Mode
Norm
m-m
Period (PER)
60
ns
---
Width (WID)
IO ns
10
ns
Leading Edge (LEE)
1 ns
1
ns
Trailing Edge (TRE)
1
ns
1
ns
High Level (HIL)
3.2 V
3.2
V
Low Level (LOL)
ov
ov
Delay (DEL)
0 ns
0 ns
Output Mode
ENABLE
ENABLE
3.
Assign the pod under test to
Analyzer 1 in
the
System Configuration
menu
as in figure 3-5. Refer to steps a through c if you are unfamiliar with menus.
Unassl
c
;tkz tinal yrer
Figure
3-5.
*System Configuration
for Data Test
1
a. Move the cursor to the Type field of
Analyzer 1
and press SELECT.
b. Set the analyzer Type to
State
using the cursor and the SELECT key.
c. Move the cursor to the
Pod
to be tested and assign it to
Machine 1.
HP 1652B/1653B
Service Manual
Performance
Tests
3-7