Saving Compare Images - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Note f#
When the trace mode is repetitive and Store exception to disk is on, the
following process takes place: data is acquired until the stop criteria is
met, data aquistion will stop, data in the acquisition memory will be
stored on the disk, and data acquisition will resume when the data is
stored. This process continues until the disk is full. The data is stored in
the same file name; however, the last three characters will
automatically be replaced with a numerical serial number. For
example, EXCEPTION will change to EXCEPT001 the second time
memory is stored.
-.
You may also specify a stop measurement based on time between the X
and 0 markers in the Compare or Difference Listing menus. This is
available only when Count is set to Time in the State Trace menu. If the
Stop Measurement is set to run until Compare Equal or Compare Not
Equal in the Compare or Difference Listings, the Stop Measurement
on time X to 0 will change to run until Compare Equal or Compare
Not Equal in the other state display menus (i.e. State Listing).
Locating
The Find Difference feature dews you to easily locate any patterns
Mismatches
in
that did not match in the last comparison. Occurrences of differences
I
tne 1
m Yff erence
Listing
are found in numerical ascending order from the start of the listing.
The first occurrence of an error has the numerical value of one.
.
This feature is controlled by the Find Difference
field in the
Difference Listing menu. When you select this fi=
Integer Entry
pop-up appears in which you enter a number indicating which
difference you want to find. The listing is then scanned sequentially
until the specified occurrence is found and rolled into view.
Saving
Compare
Images
When you save a logic analyzer configuration to a disk, the compare
images for both state analyzers are saved with it. The compare data is
compacted to conserve disk space. Likewise, when you load a
configuration from disk, valid compare data will also be loaded.
State Compare Menu
12-8
HP 16528/1653B
Front-Panel Reference