Count
Note d
The Count field allows you to place tags on states so you can count
them. Counting cuts the acquisition memory in half from lk to 512 and
the maximum clock rate is reduced to 16.67 MHZ.
Count (State Trace menu) is turned off when "Clock Period" is set to
< 60 ns in the State Format Specification menu since the clock rate is
greater than 16.67 MHz. If you select Count, the clock period
automatically changes to > 60 ns.
Selecting this field gives you the following pop-up menu.
Count
Off
Et
Time
States
Figure 10-24. Count Pop-Up Menu
Off
If you select Off, the states are not counted in the next measurement.
Time
If you select Time counting, the time between stored states is measured
and displayed (after the next run) in the State Listing under the label
Time. The time displayed can be either relative to the previous state or
to the trigger. The maximum time between states is 48 hours.
An example of a state listing with time tagging relative to the previous
state is shown in figure
10-25.
State Trace Menu
lo-18
HP 16528/1653B
Front-Panel Reference