5. Press the TRACE/TRIG
key and set the Mode to Immediate
and the
Run mode to Repetitive as in figure
3-51.
IscopSl-
Trigger
ilode
1 Immediate
1
Run mode
I-
s/Dlv
~1o.oo]
Delay
I
Armed
by
I T
r
CHl
J
:
: .
Figure 3-51. Trigger Men&
6. Press DISPLAY
and set the Markers to Time, Display to AVG# 32, Connect
dots On, and Grid On as in figure 3-52.
I[Scope}-
ldnvef ornks
llarkers
[
Time
]
Display
1-1
Connect
dots ['Z-j
x to 0 [O]
Tr lg to X j-01
Trig
to O[o]
s/Dlv
~1o.00~
Delay
I
GritI
I
Figure
3-52.
Waveforms
Display Menu
7. Press FORMAT/CM
and set V/Div and Offset according to the frost line
of the following table.
V/Div
Offset
Test
Power Supply
Tolerance
Off set Result
1.0 v
20.0 v
20.0 v
~0.502V
19.50 - 20.50 v
500
mV
10.0 v
10.0 v
to.242
V
9.758 - 10.242 V
200
mV
5.0 v
5.0 v
to.122
v
4.878-5.122
V
100 mV
2.0 v
2.0 v
20.052
V
1.948 - 2.052 V
8. Set the power supply to the voltage listed on the fixst line of the previous
table.
HP 16528/1653B