Statistics: X to 0 marker statistics are calculated for repetitive
acquisitions. Patterns must be specified for both markers and statistics
are kept only when both patterns can be found in an acquisition.
Statistics are minimum X to
0
time, maximum X to 0 time, average X to
0 time, and ratio of valid runs to total runs.
Run/Stop Functions
Run: Starts acquisition of data in a specified trace mode.
Stop: In single trace mode or the frost run of a repetitive acquisition,
STOP halts the acquisition and displays the current acquisition data.
For subsequent runs in repetitive mode, STOP halts the acquisition of
data and does not change current display.
Data Display/Entry
Display Modes: State listing; timing waveforms; interleaved,
time-correlated listing of two state analyzers (time tagging on);
time-correlated state listing and timing waveform display (state listing
in upper half, timing waveform in lower half, and time tagging on).
Timing Waveform: Pattern readout of timing waveforms at X or 0
marker.
Bases: Binary, Octal, Decimal, Hexadecimal, ASCII (display only),
and User-defined symbols.
Specifications and Operating Characteristics
F-10