Performance
Test Interval
Periodic performance verification of the HP 1652B/1653B is required at two year
intervals. The instrument's performance should be verified after it has been
serviced, or if improper operation is suspected. Calibration
should be performed
before any performance verification tests. Further checks requiring access to the
interior of the instrument are included in the adjustment section, but are not
required for the performance verification.
Performance
Test
Procedures
Au performance tests should be performed at the instrument's environmental
operating temperature
and after a 15 minute warm up. The performance tests
for the HP 1652B/1653B are separated into two sections. The first section
contains the performance verification tests for the logic analyzer portion of the
HP 1652B/1653B and the second section contains the performance verification
tests for the oscilloscope portion.
Procedures are based on the model or part
number for the recommended
equipment.
HP 16528/1653B
Performance
Tests
Service
Manual
3-3