Preset= Scales the vertical range, offset, and trigger level to predetermined
values
for displaying ECL or TTL waveforms.
beset
Vertical Range
TT.L*
1.5 v
ECL"
500 mV
* Values when Probe = 1O:l.
Trigger Level
1.60
V
-1.3 v
Calibration:
Offset, attenuation, gain, trigger level, delay, and set to defaults.
Probe Compensation
Source: The external BNC supplies a square wave signal of
approximately-400
mV to -900 mV at approximately
1.25 kHz.
Interactive
Measurements
Acquisition
Oscilloscope, timing, and state can occur simultaneously
or in series.
Mixed
Displays
Timing channels and oscilloscope channels can be displayed on the same screen.
Multiple
state machine listings can be displayed with time tags on the same screen.
Timing channels can be displayed with a state listing with Time Tags turned on.
State listings with time tags, timing channels, and oscilloscope channels can be
displayed on the same screen.
Time
Correlation
Time Interval
Accuracy
between
Modules
All modules are time correlated with the exception of when the oscilloscope is
being armed by the logic analyzer, and when the oscilloscope is not in trigger
immediate
mode.
Equals the sum of channel to channel time interval accuracies of each machine
used for a measurement.
General
The following general characteristics for the HP 1652B/1653B include the
Characteristics
environment
operating conditions, shipping weights, and instrument dimensions.
Operating
Temperature
Environment
Instrument:
Operating:
O°C to +55"C (32°F to + 131°F).
Non-operating:
-4OOC to + 70°C (-40°F to + 158°F).
Probes and Cables: O°C to 65OC ( + 32°F to +
149°F).
Disk Media:
10°C
to 50°C (+ 50°F to + 149°F).
HP 16528/1653B
Service
Manual
General
Information
1-19