VOltage
Description:
Measurement
Accuracy
Test
This test verifies the voltage measurement accuracy of the instrument.
Specification:
t (Gain Accuracy + Offset Accuracy + ADC Resolution)
Equipment
Required:
Power Supply
. . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
6114A
Digital Multimeter
. . . . . . . . . . . . . . . . . . . . . . . . . HP
3478A
BNC Cable
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . HP
10503A
BNC (f)-to-Banana
(m) Adapter
. . . . . . . . . . . . . HP
1251-2277
Banana (m)-to-Banana
(m) Cable
(2)
. . . . . . . . . HP
11000-60001
Procedure:
1. Connect the HP 1652B/1653B and test equipment as in figure
3-43.
l-
INPUT
MULTIMETER
O *-l-l
\
POWER
SUPPLY
-
&
/
LOGIC
ANALYZER
I
01652E14
Figure
3-43.
Setup for Voltage
Measurement
Accuracy
2. Connect the power supply to Channel 1 of the HP 1652B/1653B oscilloscope.
3.
In the
System Configuration
menu, turn both State/Timing
Analyzers off and
turn the oscilloscope on as shown in the previous test figure
3-42.
HP 16528/16538
Performance
Tests
Service Manual
3-37