Marker Measurements - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Marker
In addition to automatic parametric measurements, the oscilloscope
Measurements
also has two markers for making time and voltage measurements either
--
manually (Time) or automatically (Search).
To demonstrate how to make marker measurements, connect the
oscilloscope as shown in figure 25-3 and reset the oscilloscope as
described below.
Turn the instrument on.
From the System Configuration menu, turn the oscilloscope on
and all analyzers off.
Touch CHAN key, then from the default Channel Specification
menu, make the following changes:
Waveforms Menu
25-12
Input CH
1
Probe field to 1:l
Input CH 1 Impedance field to 50 Ohms
Input CH 2 Probe field to 1:l
Input CH 2 Impedance field to 50 Ohms
4. Select Autoscale, then select Continue.
HP 16528/1653B
Front-Panel Reference