The sample rate varies from 20 Hz to 50 MHz (50 MS/sample to
20 &sample) and is automatically selected by the timing analyzer to
insure complete data in the window of interest.
When your timing analyzer triggers on a glitch and displays the data,
the glitch appears in the waveform display as shown below.
ANALYZER
INPUT
SAMPLE
:
4
4
4
POINTS
I
f
ANALYZER
DISPLAY
I+---
GLITCH
DISPLAYED
ON NEXT SAMPLE
I
Figure 18-5. Glitch in Timing Waveform
Label
The Label fields contain the labels that you defme in the Timing
Format Specification menu. If there are more labels than can fit on
screen, use the left/right ROLL key and the KNOB to view those that
are not displayed.
Base
The Base fields allow you to specify the numeric base in which you
want to define a pattern for a label. The Base fields also let you use a
symbol that was specified in the Timing Symbol Table for the pattern.
Each label has its own base defmed separately from the other labels. If
you select one of the Base fields, you will see the following pop-up
menu. Decide which base you want to define your pattern in and select
that option.
Figure 18-6. Base Pop-up Menu
Timing Trace Specification
Menu
HP 16528/1653B
18-6
Front-Panel Reference