24. Connect a BNC cable from the Probe Compensation
output on the rear
panel to the channel 1 input of the HP 1652B/1653B oscilloscope.
The
instrument will prompt you when you need to switch to the channel 2 input.
ale
I
Note
a
If you use a 1O:l probe in place of the recommended
1:l BNC cable, use the
BNC-to-mini
probe adapter supplied with the instrument.
Then set the
attenuation field in step 25 to 1O:l.
25. Set the attenuation field in the calibration menu to the appropriate setting.
26. To proceed with the calibration, select Continue using the front-panel knob
and SELECT key.
27. When the calibration is complete, the updated calibration status appears on
screen and the instrument remains in the Calibration
menu.
Caution
Do not execute Set to Default after calibrating the instrument.
Otherwise, your
calibration factors will be replaced by default calibration factors.
28. Select Done with the front-panel knob and SELECT key to exit the
Calibration
menu.
-
Adjustments
and Calibration
HP 1652B/1653B
414
Service
Manual