To Perform The Self-Tests - HP 1660 Series Service Manual

Logic analyzers
Hide thumbs Also See for 1660 Series:
Table of Contents

Advertisement

To perform the self-tests

The self-tests verify the correct operation of the logic analyzer. Self-tests can be
performed all at once or one at a time. While testing the performance of the logic
analyzer, run the self-tests all at once.
The performance verification ( PV) self-tests consist of system PV tests, analyzer PV
tests, and oscilloscope PV tests ( HP 1660AS series only) .
Disconnect all inputs, then turn on the power switch. Wait until the power-up tests
1 1
are complete.
Press the System key. Select the field next to System, then select Test in the pop-up
2 2
menu.
Insert the disk containing the performance verification (self-tests) into the disk
3 3
drive. Select the box labeled Load Test System, then select Continue.
Press the System key. Select the field next to Sys PV, then select System Test in the
4 4
pop-up menu.
3–3