HP 1652B Getting Started Manual page 497

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Error Messages
c-4
Slow or missing Clock. Indicates the state analyzer has not recognized
a clock for 100 ms. Check for a missing clock if the intended clock is
faster than
100
ms. If clock is present but is slower than 100 ms, the
data will still be acquired when a clock is recognized and should be
valid.
Specified inverse assembler not found. Indicates
specified cannot be found on the disk.
the inverse assembler
State clock violates overdrive specification. Indicates the data from the
last acquisition is not reliable due to the state clock signal not being
reliable. Check the clock threshold for proper setting and the probes
for proper grounding.
States Remaining to Post Store. In dicates the nu mber of
required until memory is filled and acquisi tion is complet
states
:e.
Time count need to be turned on. This message appears when the logic
analyzer attempts to time correlate data and "Count" is not set to
"Time."
Transitions Remaining to Post Store. Indicates the number of
transitions required until memory is filled and acquisition is corn plete.
Unsupported destination format-file not copied. Indicates the disk you
have attempted to copy to is either not formatted or formatted in a
format not used by the logic analyzer. Format the disk or use a properly
formatted disk and repeat the copy operation.
Value out of range. Set to limit. Indicates an attempt to enter a value
that is out of range for the specific variable. The logic analyzer will set
the value to the limit of the variable range automatically.
Waiting for Arm. Indicates the arming condition has not occurred.
Waiting for Prestore. Indicates the prestore condition has not occurred.
HP 16528/1653B
Front-Panel Reference