Not In Range To - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

-
Time range fields
In range
to
When the In range-to
field is selected from the pop-up, two time
value fields appear next to the In range field. See figure 25-18. The time
range for the stop condition is entered using the keypad. When either
time value field is selected, the time value can be entered the same as
for the Less Than field. Default values for In range
to-
is
10
ns.
When In range-to-
is selected, the oscilloscope will run until the
X-O time interval is in the range of the time values entered for the In
range-to-
time fields. When the condition is met, the oscilloscope
will stop acquisitions and display Stop condition satisfied.
(Specif
u tiarkers
)
tlarkers
Ir
[
Search
]
Display
(
Normal
]
Connect
dols[r]
I
1
b
Search
tbrkers
(Done)
I I
Rarke:
Type
[]
X+arker
sel
on
1x1
at
the
pGKiT]
slope.
@Marker
set
on 11
at
the
[T]
[
50%
]
level
uith
level
ui th
I I
-1
slooe.
Stop
measurefnent
uher
X+3
[
Ir
raw9
][Ob]tO
L-1
Store
exceptlon
to
disk:
I-1
Figure
2548.
Range Fields
Not in range
to
When the Not in range-to-
field is selected from the pop-up, two
time value fields appear next to the Not In Range field. The time range
for the stop condition is entered in these time fields. When either time
value field is selected, the time can be entered the same as for the Less
Than field.
When Not In range- to
is selected, the oscilloscope will run until
the X-O time interval is not in the range of the time values entered for
the Not in range-to-
time fields. When the condition is met, the
oscilloscope will stop acquisitions and display Stop condition satisfied.
Default values for these fields are
10
ns.
HP 16528/16538
Waveforms Menu
Front-Panel Reference
25-21

Advertisement

Table of Contents
loading

This manual is also suitable for:

1653b

Table of Contents