5. Set the Input to CH 1, V/Div to 100 mV, Offset to 0 V, Probe to l:l,
Impedance to 50 Ohms, and s/Div to 5.0 us as in figure 3-55.
t c:H
1
Figure 3-55. Channel
Menu Configuration
6. Press TRACEITRIG
and set 'the Run mode to Repetitive and trigger Level to
0 V as in the previous figure 3-51.
7. Press DISPLAY
and set Markers to Time, Display to Normal, Connect dots
On,
and
Grid On as in figure 3-56.
-Scope-
liavef
arms
Auto-Measure
tlarkersl
Time
1
Display
[NDrmal)
Connect
dots
Ion]
x
to
0
[OJ
Trlg
to
X 1-1
Trig
to
0-0
s/Dlv
pmEiq
Delay
I
Grid
I
: . . . . : . . . .
Figure 3-56. Waveforms
Display Menu Configuration
8. Set the signal generator for 100 kHz at -4.5 dBm and press RUN on the
HP 1652B/1653B. The signal on screen should be five cycles at two divisions
amplitude.
9. Press TRACEITRIG
and adjust the trigger Level for a stable display. The
signal on screen should be five cycles at approximately
2 divisions amplitude.
10. Press DISPLAY
and set Display to AVG# 16.
Performance
Tests
3-46
-
HP
16528/1653B
Service Manual