Oscilloscope
The following specifications are the performance standards or limits
Specifications
against which the oscilloscope in the HP 1652B/l653B is tested.
Vertical
Bandwidth (-3 dB):
dc to 100 MHZ (single shot).
DC Gain Accuracy: t 3%
of full scale.
DC Offset Accuracy:
2 (2 mV + 2% of the channel offset + 2.5% of
full scale).
Voltage Measurement Accuracy (DC): (Gain
accuracy + ADC
resolution + Offset accuracy).
Horizontal
Time Interval Measurement Accuracy: $276 X
s/div + 0.01%
X
delta-t + 500 ps).
Trigger
Sensitivity
10% of full
screen.
Oscilloscope
The following operating characteristics are not specifications, but are
Operating
typical operating characteristics for the oscilloscope in the HP
1652B/1653B. These are included as additional information for the
Characteristics
user
.
Vertical
Transition Time (109b
to !W%):
s
3.5 fls.
(at
BNC)
Number of Channels: 2.
Vertical Sensitivity Range: 15
mV/div to 10 V/div (1: 1 probe).
Vertical Sensitivity Resolution:
Adjustable 2 digit resolution.
Maximum Sample Rate:
400 MSamples/second.
Analog-to-Digital Conversion:
6 bit real-time.