Search Marker Measurement Example; Connecting The Equipment; Making The Measurement - HP 1652B Getting Started Manual

Logic
Hide thumbs Also See for 1652B:
Table of Contents

Advertisement

Search Marker
The following example will show how to make an automatic marker
Measurement
measurement using the Search markers. We will set the markers to
Example
make an X-O marker measurement on the CH 1 and CH 2 input
waveforms. We want to measure the time between the falling edge of
the 2nd displayed pulse on CH 1 to the rising edge of the 5th displayed
pulse on CH 2. We'll perform the measurement from the 10% point on
CHl to the 90% point on CH 2.
Connecting
the Equipment
Connect the equipment as shown in figure 25-20.
I
Note m
An extra long BNC cable is used on channel 2 so the signal is delayed.
Waveforms
Menu
25-24
BACK
FRONT
VIEW
VIEW
EXTRA
LONG CABLE
Figure 25-20. Equipment Setup For
Search Marker Example
Making the Measurement
1. In the Waveforms menu, select Markers field. When the pop-up
appears, select the Search field. With Markers set to Search, two
new fields will appear: X-pattern
from start, and O-pattern
from start. Set the X-pattern to 2, and the O-pattern to 5. Set the
s/Div to 500~s. See figure 25-21.
HP 1652B/1653B
Front-Panel Reference