Performance
Tests
HP 16528/1653B
3-14
Service
Manual
7. Connect the next clock to the test connector and repeat steps 4 and 6 for the
appropriate
clock. Repeat these steps until all clocks have been tested
(clocks J, K, L, M and N).
8. Remove the probe tip assembly from the logic analyzer probe cable and
attach it to the next logic analyzer probe cable to be tested. Take care not to
dislodge grabbers from the test connector.
9. Repeat steps 3,4,6, and 7 until the lower bits of all pods (pods 1 through 5)
have been tested with all clocks.
10. Disconnect the lower eight bits (bits 0 through 7) from the test connector.
Attach the upper eight bits (bits 8 through 15) to the test connector and
repeat steps 3,4,6,7,
and 8 until the upper bits of all pods (pods 1 through 5)
have been tested with all clocks.