5 Operation

5.4 Adapting basic settings

+
26
5.4
Adapting basic settings
Adapting basic settings
Exposure times for the use of films with the sensitivity class E are factory
pre-set, as well as the XIOS/XIOS
IMPORTANT
The exposure times for sensor and film images are programmed
separately. The factory pre-set sensor programming is configured for
XIOS/XIOS
The basic setting must be adjusted for other exposure conditions.
Deviating exposure conditions:
E for films of sensitivity class E such as Kodak Ekta Speed, Agfa-Dentus
M2
D for films of sensitivity class D such as Kodak Ultra Speed
Set the exposure time for films of sensitivity class D three levels higher
with the plus button.
Film and development conditions
Varying film and development conditions can result in additional
deviations of one time level up or down.
IMPORTANT
The configuration of the film and sensor keys permits flexible adjustment
to various film sensitivity classes and sensors. It is also possible to set
up the exposure adjustment for another film sensitivity class via the
sensor key, if no sensor is being used.
Reprogramming the base values
+
1. Press the film, sensor and plus buttons at the same time.
Service S01 appears in the display.
+
+
2. Use the plus or minus key to select the base value that is to be
changed:
S01 corresponds to the base value for film
_
S02 corresponds to the base value for sensor
S03 corresponds to the software version
3. Confirm the entry with the Film key.
The current base value for film or sensor is now displayed:
Plus
Plus
sensors.
Sirona Dental Systems GmbH
Operating Instructions HELIODENT
sensors.
62 14 717 D3507
D3507.201.01.06.02 05.2011