8 Maintenance

8.1 Care, cleaning, disinfection and sterilization

50
8
Maintenance
WARNING
Danger of touching live parts
If the housing is damaged, there is a possibility of touching live parts
inside the unit. If the housing is damaged, the unit must be put and
left out of operation until it has been professionally repaired.
NOTE
Regular inspection
Some countries have legal regulations which require regular safety
inspections of electrical devices or systems by the operator.
Sirona would like to draw your attention to the fact that a so-called
"retest" (repeat test) must be carried out for the CEREC AC
acquisition unit every three years at the latest. In addition, this retest
also must be performed following every repair or retrofit of
components such as the PC, the PC power supply, the isolating
transformer, the CEREC Bluecam and the camera cable.
NOTE
Annual maintenance performed by trained technical personnel is
recommended.
Care, cleaning, disinfection and sterilization
8.1
NOTE
Use only chemical products recommended by Sirona.
8.1.1

Cleaning and disinfection agents

Kerr Corporation
Patterson
> 60% isopropyl alcohol
8.1.2

Monitor screen

Cleaning
NOTE
Never spray the monitor screen with a disinfectant or cleaning spray.
The monitor screen must be wiped off with a soft cloth.
Sirona Dental Systems GmbH
Operating Instructions CEREC AC
● CaviCide
● Cavi Wipes
● pdCARE
● pdCARE Wipes