Care and maintenance

Care and maintenance
5.1

Dismantling SIROInspect

Dismantling SIROInspect
The SIROInspect can be dismantled for cleaning, disinfection or
sterilization.
NOTICE
Risk of short circuit
D
The battery (D) and the base body (C) must not under any
circumstances be immersed in liquid. These components can only be
disinfected by wiping them down.
1. Pull the battery (D) out of the base body (C).
2. Unscrew the light pen (A or B) from the base body.
C
3. Slide the attachable filter (E) off the base body.
4. Carry out cleaning, disinfection, and sterilization.
5.2

Cleaning the outer surface

Cleaning the outer surface
Combining cleaning and disinfection
Manual cleaning must always be combined with disinfection.
Cleaning SIROInspect
The SIROInspect must not be cleaned or disinfected with a machine.
1. Remove the silicone ring from the attachable filter.
2. Clean the light guide, the silicone ring and attachable filter by placing
it under running water (< 38 °C, water must be at least drinking water
quality).
3. Wipe down the battery and base body with a damp cloth. NOTICE!
The battery and base body must not be cleaned under running water.
NOTICE
Never clean in an ultrasonic bath!
5.3

Disinfecting the surface

Disinfecting the surface
Immersion in disinfectants, SIROInspect
The SIROInspect must not be cleaned or disinfected with a machine.
SIROInspect can be disinfected by wiping it. NOTICE! The battery and
base body must not be disinfected by immersion.
NOTICE
Never immerse in disinfectants!
5 Care and maintenance