6.12 Replacing the bite block holder
A
Removing a defective bite block holder
Attaching a bite block holder
6 – 38
6.12
Replacing the bite block holder
6.12.1
For units without an occlusal bite block
1.
Loosen the two screws A and remove the bite block holder.
2.
Replace silicone rings, see chapter 6.12.3
Place the bite block holder into the groove of the silicone rings on the tube
bend and fasten it with the two screws A.
ATTENTION: What to do after replacement?
– Check the system adjustment via a diaphragm test exposure
PAN - Symmetry (see page 7-25).
– Perform a complete PAN unit adjustment (see chapter 4).
Tab 6
59 38 399 D3352
D3352.076.01.26.02 09.2016