Site Requirements and Specifications

Performance Specifications

Performance Specifications

Table 4
Type
Temperature range
Temperature stability
Temperature accuracy
Column capacity
Warm-up/cool-down
time
Dead volume
Communications
Safety and
maintenance
GLP features
Housing
All specifications are valid for distilled water at ambient temperature (25 °C), set point at
N O T E
40 °C and a flow range from 0.2–5 ml/min.
28
Performance Specifications Thermostatted Column Compartment
Specification
10 degrees below ambient to 100 °C
up to 80 °C: flow rates up to 5 ml/min
up to 100 °C: flow rates up to 2.5 ml/min
± 0.05 °C
± 0.8 °C
± 0.5 °C
Three 30 cm
5 min from ambient to 40 °C
10 min from 40 – 20 °C
1.6 µl low dispersion heat exchanger
3 µl left heat exchanger
6 µl right heat exchanger
Controller-area network (CAN), RS-232C,
APG Remote: ready, start, stop and
shut-down signals, LAN via other
modules
Extensive support for troubleshooting
and maintenance is provided by the
Instant Pilot, Agilent Lab Advisor, and the
Chromatography Data System.
Safety-related features are leak
detection, safe leak handling, leak output
signal for shutdown of pumping system,
and low voltages in major maintenance
areas.
Column-identification module for GLP
documentation of column type
All materials recyclable.
Comments
With calibration
i.d. 0.12 mm (capillary kit
available)
i.d. 0.17 mm (standard)
1290 Infinity TCC User Manual