Performing a Fixed Kinetics Measurement
Performing a Scan Kinetics Measurement
Related Topics
Performing a Scan Kinetics Measurement
Data Display
Performing a Fixed Kinetics Measurement
Kinetics measurements can be made at one wavelength or at more than one wavelength
simultaneously. This includes multicell kinetics, which allows running a fixed-wavelength
measurement on several samples simultaneously using a sample changer.
To perform a fixed kinetics measurement
1. Make sure Single wavelength or Multiple wavelength is selected on the
Settings in Rate.
2. Click Measure Rate.
3. Click Zero.
4. In cell position 1 load the sample to use to measure the zero and choose OK.
5. Click Measure.
If necessary, the Zero action button can be used to remeasure the zero sample.
6. When samples information appears, modify it if desired.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.