Maintenance
Cleaning the Q00 Ion Optics

Shutting Down and Venting the System

Shut down and vent the system as described in the topic
on
page 48

Removing the Ion MAX-S Ion Source Housing

Remove the Ion MAX-S ion source housing to access the ion source interface assembly. Remove
the Ion MAX-S ion source housing as described in the topic
MAX-S Ion Source Housing"

Removing the Ion Source Interface Assembly

Remove the ion source interface assembly to access the Q00 ion optics assembly. Remove the ion
source interface assembly as described in the topic
page

Removing the Ion Optics Cage Assembly

1. Using a Phillips screwdriver, remove the three screws fastening the ion source interface cover to
2. Using a hex ball driver, remove the roughing pump adapter access plate. See
80
TSQ Series Hardware Manual
.
73.
To remove the ion optics cage assembly
the mass spectrometer, and place the cover in a safe place. See
Figure 47. Removing the ion source interface cover screws
Ion source
interface cover
screws
on
page
63.
"Removing the Ion Source Interface"
"Shutting Down the System Completely"
"Removing the Ion MAX or Ion
Figure
47.
Figure
on
48.