Appendix B

4000 Series

4000-
Memory Allocation Error.
Cause:
Action:
4002-
Thermal Transpiration correction had no effect.
Cause:
Action:
4003-
Error Converting Pressures.
4004-
Error Computing Volume Adsorbed.
Cause:
Action:
4005-
Pressures were not smoothed. Not enough pressures below 0.10 P/Po.
Cause:
Action:
202-42811-01 - Mar 2011
An internal processing and/or hardware error occurred during report
generation.
Contact your service representative if you continue to receive this error
message.
The Thermal transpiration correction option was selected on the Report
Options dialog box. However, the correction did not change any pres-
sure by more than one percent.
De-select this option to disable this message. This correction is only
meaningful for very low pressures.
An internal processing and/or hardware error occurred during report
generation.
Contact your service representative if you continue to receive this error
message.
The Smooth pressures below 0.10 P/Po option was selected on the
Report Options dialog. There must be at least 10 pressures within this
range for smoothing to occur.
Deselect this option to disable this message.
ASAP 2020 Confirm
B-25