ASAP 2020 Confirm
2475- (Unit {number}) Communications port COM (port number) specified in the pro-
gram control files cannot be accessed. Unit cannot initialize.
Cause:
Action:
2476- Unit startup failed. Cannot initialize the communications port.
Cause:
Action:
2477-
(Unit n; Serial nn) did not properly initialize.
Cause:
Action A:
Action B:
Action C:
2478- Error copying sequential data segment.
Cause:
Action:
B-12
The operating system has prevented the application from accessing the
communications port.
Review the hardware configuration of the computer, ensure that no
other application is using the port. Contact your service representative
if you continue to receive this error message.
The communications port specified during installation is invalid.
Use the setup program to change the analyzer configuration. Contact
your service representative if you continue to receive this error mes-
sage.
The software was unable to initialize the analyzer.
Run the Setup program and ensure that a valid port is specified; if not,
specify a valid one when prompted.
Reinstall the software, then restart application.
Contact your Micromeritics service representative if you continue to
get this message.
An internal processing and/or hardware error occurred while accessing
a portion of a sample file.
Confirm that the media being accessed does not contain errors; for
example, you may wish to use a utility such as ScanDisk. Contact your
service representative if you continue to receive this error message.