(Unit [number]) Analysis suspended: Time limit exceeded while dosing mani-
fold to (target pressure point) mmHg with (gas name).
Cause A:
Action A:
Cause B:
Action B:
6528-
Low pressure gauge offset too high - [PR4] [PR-U].
Cause:
Action:
6529-
Master pressure gauge offset is too high - [PR1] [PR-U].
Cause:
Action:
6530-
Volume calibration canceled due to failure (code [0]).
Cause:
Action:
B-54
The maximum time was exceeded before the target pressure point was
reached. The nitrogen regulator may be set too low or turned off.
Set the analysis gas regulator to 10 psig (0.7 bar). Then resume the
analysis.
The analysis gas bottle is empty.
Connect a new analysis gas bottle. Then resume the analysis.
A check of the 1000 mmHg gauge's offset during an automatic oper-
ation indicated it was too high.
If this message occurs repeatedly, contact a Micromeritics service rep-
resentative
A check of the 10 mmHg or 1 mmHg gauge's offset during an auto-
matic operation indicated it was too high.
If this message occurs repeatedly contact a Micromeritics service rep-
resentative.
A problem occurred during volume calibration.
Contact a Micromeritics service representative.
Appendix B
202-42811-01 - Mar 2011