Micromeritics ASAP 2020 Confirm Operator's Manual page 210

Developer/analyst
Table of Contents

Advertisement

ASAP 2020 Confirm
Reference Volume
Reference Volume Temp. Enter the temperature of the reference volume. Since the ref-
Perform measurement
Start
5-24
Enter the volume in cubic centimeters for the reference vol-
ume you are using in this calibration.
erence volume should be submerged in an ice bath, this
value should remain at 0.0 for best accuracy.
Choose the number of measurements you wish to perform;
you may choose up to 9.
Begins the calibration; a status view is displayed. After the
calibration has finished, a confirmation view of the Calibrate
System Volume dialog is displayed.
This dialog allows you to choose the volume you wish to
use.
Previous
The system volume and lower manifold volume in use
before this measurement.
Entered
An entered measurement.
Measured
The last measured system and lower manifold volumes; for
example if you choose 3 measurements, this value is that of
the last (third) measurement.
Average
The average of the series of measurements; for example if
you choose 3 measurements, this value is the average of the
three measurements.
Click
to close the dialog.
Done
Calibration
202-42811-01 - Mar 2011