Preventive Maintenance
8. Check the washer or O-ring used at the oil-filling port; replace if necessary. Replace the
oil-fill plug.
9. Place the vacuum pump back into the analyzer onto the drip tray and reconnect the
vacuum pump hose.
10. Reconnect the vacuum pump power cord.
11. Replace the vacuum pump panel.
12. Place the analyzer ON/OFF switch in the ON position. The pump must run a few hours (or
overnight) to eliminate air and moisture from the fresh fluid and to produce efficient
vacuum operations.
202-42811-01 - Mar 2011
ASAP 2020 Confirm
8-11