Appendix B
6200 Series
6201-
'GetOpenFile()' reports error code (number). File selection cannot be made.
Cause:
Action:
6202-
Invalid communications port (port name) specified in the application's config-
uration settings. Unit cannot initialize.
Cause A:
Action A:
Cause B:
Action B:
6203-
No serial number available. Unit cannot initialize.
Cause:
Action:
6204-
The disk does not have enough room for the ROA data. The space required is
(number) megabytes.
Cause:
Action:
202-42811-01 - Mar 2011
You specified a file name which does not exist.
Confirm that the file name used is valid and that the file exists.
The communications port assigned to the ASAP 2020 cannot be made
available for proper communications.
Run the ASAP 2020 Setup program, select Change analyzer setup, and
enter the correct port information. If the message continues, contact
your service representative.
The ASAP 2020 control file was manually modified and an invalid
communications port was specified.
Use the ASAP 2020 Setup program to change the communications port
assigned to the ASAP 2020.
The ASAP 2020 control file does not contain a serial number for the
unit.
Run the ASAP 2020 Setup program, select Change analyzer setup, and
define the serial number for the instrument.
The computer's disk is almost out of disk space.
Delete unwanted files.
ASAP 2020 Confirm
B-43